About Vertical flame test chamber
Parameter
| Item |
Parameter |
| Inner size |
About 330*330*765mm. |
| Windows |
tempered glass observation window |
| Exhaust holes |
16pcs proof holes,6pcs each side(total 12pcs) |
| Exhaust diameter |
12.5mm |
| Proof hanging clamp |
located 17mm above the highest point of the igniter nozzle. |
| Diameter of the igniter nozzle |
11mm |
| Angle about the head and vertical line |
25 |
| inlet gas pressure of the igniter |
17.2kpa1.7kpa |
| Ignition time |
accurate to 0.05s |
| Gas source |
industrial grade propane, butane or propane / butane mixed gas. When the sample is pretreated according to 7.3 in GB5455-2014. methane gas with a purity of not less than 97% is selected. |
| Sample size |
300*89mm, five horizontal and vertical |
| Ambient temperature |
-10 C 30 C |
| Power supply voltage |
220V10% 50HZ |
| Power |
less than 100W |
| Flame time |
0-99M99S |
| Flame time accuracy |
0.01s |
| Afterglow time |
longest 0-99M99S |
| Blue flame height |
15mm-175mm with flame height scale |
| Ignition time |
any setting (can be preset) |
Features
Gas pipeline system special designed, low space occupied, and convenience to move;
Exhaust vents at the bottom and top of test chamber ensure good ventilation and steady flame;
LED displayer shows the igniting time, afterflame time and afterglow time, with high working efficient and the data more accuracy;
A stainless steel construction, large glass-viewing window;
Excellent reliability and minimum maintenance;
Purpose-designed test chamber.
Advanced Safety and Control FeaturesThis vertical flame chamber incorporates essential safety elements such as a gas leak detector and an emergency stop mechanism to protect users during operation. The intuitive touch membrane control panel simplifies test setup, while electronic piezo ignition ensures rapid, fail-safe flame initiation every time. Its robust construction using stainless steel and powder coated mild steel guarantees durability and compliance with international testing standards.
Flexible and Precise Testing CapabilitiesEngineered to test a range of plastic, rubber, and composite samples, this chamber accommodates up to three specimens in a single batch. Flame height adjusts from 20 mm to 175 mm to suit various standards, with digital timing and gas pressure control for test accuracy. The observation window and removable exhaust chimney enhance user experience by providing visibility and proper ventilation during tests.
FAQ's of Vertical flame test chamber:
Q: How do I operate the vertical flame test chamber for plastic material testing?
A: To operate the chamber, place your prepared specimen in the supplied holder and align it with the sample tools. Set the desired flame height and timer using the touch membrane control panel. Ignite the burner with the electronic piezo ignition, and monitor the test via the tempered glass window. Follow the specified standard (ASTM D5025, UL94, or IEC 60695-11-10) for accurate results.
Q: What materials and specimen sizes can be tested in this chamber?
A: The chamber is designed for plastic, rubber, and composite materials. It accommodates specimen sizes of 125 mm x 13 mm x 3 mm, adhering to popular testing standards. Up to three specimens can be tested per batch, making it suitable for laboratory and quality analysis applications.
Q: When should the removable exhaust chimney be used during testing?
A: The exhaust chimney is essential whenever testing produces fumes or emissions, as it aids in ventilating the chamber. Its removable design allows for easy cleaning and adapts to different laboratory setups, ensuring the testing environment remains safe and compliant.
Q: Where is the chamber commonly used, and who benefits from its features?
A: This flame test chamber is widely used in testing laboratories, manufacturing quality control departments, and research institutions. Material scientists, quality inspectors, and compliance officers benefit from its accuracy, safety features, and ease of use.
Q: What is the process for adjusting the flame and gas pressure?
A: Flame height can be adjusted between 20 mm and 175 mm via the control panel. Gas pressure is manually regulated using the chamber's 8 mm gas inlet, with pressure adjustable between 10-100 kPa to align with specific testing requirements.
Q: How does the gas leak detector ensure user safety?
A: The integrated gas leak detector continuously monitors the chamber for potential gas leaks during operation. If a leak is detected, it alerts the user and enables prompt protective measures, minimizing the risk of accidents.
Q: What benefits do users gain from the chamber's digital timer and observation window?
A: The digital timer (0-999 seconds) allows precise control over exposure times, essential for test reproducibility. The tempered glass observation window enables safe and clear monitoring of the specimen throughout the testing process, enhancing both safety and data collection efficiency.